Transmitter

Use

It offers solutions in the following industries:

  • Oil
  • Chemicals
  • Electricity
  • Metallurgy
  • Water
  • Brewery etc.
[/fusion_builder_column_inner]

Characteristics

The Transmitter SPE-P400 is high precision with a silicon pressure sensor designed to achieve field device communication with the help of a microprocessor. It is distinguished by its high precision and stability and is ideal for fluid, gas, steam, density and fluid pressure measurement.

PERFORMANCE SPECIFICATIONS

Reference Accuracy of Calibrated Span: ±0.075%,±0.1%  (If TD>10(TD=URL/SPAN), ±(0.005×TD)%)

(includes terminal-based linearity, hysteresis, and repeatability)

The square root accuracy is 1.5 times of reference accuracy of calibrated span.

Ambient Temperature Effects -20℃~65℃:±(0.075×TD+0.025)%×Span

-40℃~-20℃ &  65℃~85℃:±(0.1×TD+0.025)%×Span

Every 10℃ is ±0.04% ×Span (TD=1)

Static Pressure Effects ±(0.025%URL+0.05%Span)/10MPa
Over pressure Effects ±0.05%×Span /10MPa
Stability ±0.1%×Span /3 years
Power Supply Effects ±0.001% /10V (12~42V DC)
Output Two wire 4~20 mADC output with digital communications, linear or square root programmable.

HART FSK protocol are superimposed on the 4~20 mADC signal. Output range: 3.9 mA to 20.5 mA.

Failure Alarm (the mode can be selected): Low Mode (min): 3.7 mA        High Mode (max): 21 mA

No Mode (hold): Keep the effective value before the fault. Note: The standard setting of failure alarm is High Mode.

Response Time: The amplifier damping constant is 0.1 sec; The sensor damping constant is 0.1~1.6 sec, it depends on the range and range compression ratio. Amplifier damping time constant is adjustable from 0.1 to 60 sec by software and added to response time.
UpTime:  <15s
Ambient Temperature Limits: -40 to 85℃ / -20 to 65℃ with LCD display or fluorine rubber sealing
Storage and Transportation 

Temperature Limits:

-50 to 85℃ / -40 to 85℃ with LCD display
Working Pressure Limits (Silicone oil): Maximum working pressure:16MPa,25MPa,40MPa
Static Pressure Limits: 3.5kPa abs. to maximum working pressure
One-way Overload Pressure Limit: The maximum one-way overload pressure is maximum working pressure

 

PHYSICAL SPECIFICATIONS

Wetted Parts Materials

 Sensor Body  316L stainless steel
 Isolating Diaphragm  316L stainless steel / Hastelloy  C/Gold plated on 316L/FEP

plated on 316L/Tantalum

 Cover Flange  316 stainless steel
 Nuts and Bolts  304 stainless steel
 Process Connector  316 stainless steel
 Fill fluid  Silicone oil/Fluorinated oil
 Process Connector Gasket  Perbunan(NBR)/Viton  (FKM)/Teflon(PTFE)
 Amplifier Housing  Aluminum with epoxy resin coat
 Housing Gasket  Perbunan(NBR)
 Name plate and tag  304 stainless steel
 Weight  3.3kg
 Degrees of Protection:   IP67

Installation

Installation – Requirements – Information

Supply 24VDC, R≤ (Us-12V) / Imax kΩ, Imax = 23 mA. Max voltage: 42VDC, Minimum voltage: 12VDC, 15VDC (with LCD) 230Ω to 600Ω for digital communication

Connection:

An electrical connection is made via cable entry M20x1.5. The screws are suitable for cable cross-sections up to 2.5mm².

Use of 7/16-20 UNF threaded gasket and 1/4-18 NPT female thread on both sides.